2012/02/05

1448 Filled LN2 cold trap. Still using LN2 from clean room dewar.
1544 Begin LHe transfer (25.5%).
1625 End LHe transfer (73.8%)
1700 Tower temperature, although a little erratic and has been increasing since last night, rose to just over 80 mK. That is acceptable for this ionisation-only run. We should keep monitoring the tower temperature, but for now I do not believe it necessary to clean the helium cold trap. We would have to pump out the mixture, isolate the trap and evacuate it overnight. We should do this before our next run, but for now, let's just keep an eye on it.
1750 Start flashing detector.
1900 Stop flashing detector.
2145 Temperature of 1 K pot is a little cold. Open needle valve a little more.
Start taking Ba-133 data: four half-hour runs per bias: -7 V; -8 V; -9 V; and -1 V.