0900 DAQ crashed overnight. Collected all -4 V Ba-133 data, but nothing else.
1148 Filled LN2 cold trap.
1200 Start flashing detector.
1500 Stop flashing detector.
1630 Start taking Ba-133 data at -4 V, using 40 ADC bins trigger threshold.
1730 HV DCRC interface board wired up. Conduct electrical check: grounding good; and everything else looks fine:
Ebox Pin Resistor Resistance from pin to R38 input [MΩ] DCRC Pin Continuity?
1 R37 9.97 - -
2 R34 9.98 - -
3 R31 9.99 - -
4 R28 9.96 - -
5 R25 9.97 - -
6 R22 9.96 - -
7 R19 9.99 - -
8 R16 9.99 - -
9 R13 9.98 - -
10 R12 9.99 - -
11 - - 11 yes
12 - - 12 yes
13 R07 9.99 - -
14 R06 9.98 - -
15 - - 15 yes
16 - - 16 yes
17 R01 9.98 - -
18 R35 9.97 - -
19 R32 10.01 - -
20 R29 9.96 - -
21 R26 9.92 - -
22 R23 9.97 - -
23 R20 9.96 - -
24 R17 9.96 - -
25 R14 9.98 - -
26 - - 26 yes
27 - - 27 yes
28 R09 9.96 - -
29 R06 9.99 - -
30 - - 30 yes
31 - - 31 yes
32 R03 9.95 - -
33 R02 10.03 - -
34 R36 9.96 - -
35 R33 9.95 - -
36 R30 9.95 - -
37 R27 9.98 - -
38 R24 9.96 - -
39 R21 9.95 - -
40 R18 9.98 - -
41 R15 9.97 - -
42 - - 42 yes
43 R11 9.97 - -
44 R10 9.97 - -
45 - - 45 yes
46 - - 46 yes
47 R05 9.98 - -
48 R04 9.95 - -
49 - - 49 yes
50 - - 50 yes
1845 Stop data run. Three data series collected.
1900 Start helium transfer. Helium liquid level 10.0%.
1905 Start flashing detector.
1940 Stop flashing detector.
1945 End LHe transfer (92%)
1950 Stop helium transfer. Helium liquid level 92.0%.
2039 Started data taking at -10V, 40 ADC bins, 120 min series (no Ba-133) source.
Checked out the safety DCRC interface board. Looks good.
2045 Filled LN2 cold trap.