2012/01/29

1523 Filled LN2 cold trap.
1530 Noticed that 1K Pot and Tower temperature were quite high (~ 2 K). G1 and G2 were fluctuating heavily (going from 400 mbar to < 200 mbar, then up to 800 mbar, etc).
Looking through temperature history today, found that at 1330 temperature increased from 500 mk to 1.8 K
Seems that a lot of mix isn't in the fridge - high pressure in P1.
Stopped circulation. Will recondense mixture.
Closed V6.
Waited for pressures in G1/G2 to lower < 100 mbar - opened V9 to pull out mix that was in dump.
When G1 and G2 are ~ 70-80 mbar, most of mix is probably condensed
Closed V9
Opened V6 slowly. Note that P1 has dropped- most of mix is no longer outside of fridge
Pulled out more mix from dump - opened V14 (P1 increased again)
Monitored G1/G2 to make sure they were < 200 mbar. Opened/closed V6 accordingly
Note: Auto-circulation VI was still on when detectors were flashed. This should be turned off once circulation has fully commenced because this VI monitors pressures in G1/G2 and closes/opens V6 to maintain pressures < 200 mbar. Flashing detectors intereferes with this.
1650 Opened V6 fully.
1730 G1: 82 mbar , G2: 80 mbar , Tower: 128 mK
1810 Begin LHe transfer (~8%).
1840 End LHe transfer (79.7%).
G1: 75.5 mbar , G2: 71.6 mbar, Tower: 79.2 mK
2010 Tower temperature is very eratic. It has risen to 115 mK, and is still going up.
Switch to mixing chamber thermometer, use H35 curve. Mixing chamber is at 92 mK. Doesn't look like a themometetry problem.
Monitor mixing chamber temperature overnight.