2012/01/27
| 0110 | Pump out He-4 exchange gas at 4.5 K. A little late… temperature dropped very rapidly after 8 K, and was caught sleeping. Hopefully it'll be alright. Leave IVC to pump out overnight. |
|---|---|
| 1115 | Filled LN2 cold trap. |
| Temp: 5.3 K ; IVC pressure: 4E-7 mbar | |
| 1820 | Begun LHe transfer (21.4%) |
| 1851 | Tower: 5.8 K, Still: 2700 Ω, Cold plate: 2741 Ω, MC: 2685 Ω |
| 1930 | End LHe transfer (~92%) |
| Pump mixture into dump | |
| 2210 | Conduct electrical check at 4.2 K: no problems found. |
| Detector: | G102 DIB1 | BH: | Stripline: | 19 | ||||||||
| SC: | ZB1-4-8 | 300 K | LN | LHe | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2012-01-24 | Date | 2012-01-26 | Date | 2012-01-27 | |||||
| SQUET: | ZC23 | Who | HC | Who | RR & SM | Who | HC | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis | |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | ||||||||||
| Nominal: | Nominal: | Nominal: | ||||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | ||||||||||
| Phonon | ||||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.62 | OK | 80 - 140 | 104 | OK | |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2 | OK | 80 - 140 | 91 | OK | |||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.7 | OK | 80 - 140 | 101 | OK | |||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.6 | OK | 80 - 140 | 91 | OK | |||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18 | OK | 9 - 13 k | 7.3 | OK | 1.0 - 1.3 k | 1.28 | OK | |
| Feedback B/Ret | 2 | 18 | 20 | OK | 8.2 | OK | 1.0 - 1.3 k | 1.28 | OK | |||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.6 | OK | 1.0 - 1.3 k | 1.27 | OK | |||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7.1 | OK | 1.0 - 1.3 k | 1.26 | OK | |||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 107,93,87,90,88 | OK | 30 - 40 | 60.1, 52, 47, 50, 48 | OK | 20 - 30 | 47,38,32,35,33 | OK | |
| QET B/Ret | 35 | 29, 33, 47, 48 | 87,87,91,87 | OK | 47, 46, 51, 47 | OK | 20 - 30 | 32,31,36,31 | OK | |||
| QET C/Ret | 7 | 5, 9, 10, 28 | 84,85,74,84 | OK | 44, 44, 34, 44 | OK | 20 - 30 | 30,30,19,29 | OK | |||
| QET D/Ret | 38 | 36, 40, 43, 44 | 87,84,84,84 | OK | 47, 44, 44, 44 | OK | 20 - 30 | 33,30,30,30 | OK | |||
| Charge | ||||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.043 | OK | 1.0 - 1.1 k | 1.27 | OK | |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open, 39 MOhm (B) | OK | < 1 nS | open | OK | <1 nS | >55 MOhm | OK | ||
| Qo bias/QET Ret | 11 | open, 39 MOhm (B) | OK | open | OK | <1 nS | >55 MOhm | OK | ||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 44 MOhm | OK | 6 - 8 nS | 50 Mohm | OK | <1 nS | >55 MOhm | OK | |
| Qo feedback/Drain/Source | 30 | 12, 27 | 44 MOhm | OK | 50 Mohm | OK | <1 nS | 54 MOhm | OK | |||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.726 | OK | 0.9 - 1.0 V | 1.09 | OK | — | OK | ||
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.139 | – | 1.2 - 1.5 V | 1.42 | – | — | – | ||
| LED 2 | 26 | chassis | 1.128 | – | 1.41 | – | — | – | ||||
| Detector: | G102 DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-4-6 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2012-01-24 | Date | 2012-01-26 | Date | 2012-01-27 | ||||
| SQUET: | ZC70 | Who | HC | Who | RR & SM | Who | HC | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8 | OK | 1 - 2 k | 1.9 | OK | 80 - 140 | 96 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.7 | OK | 80 - 140 | 103 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.8 | OK | 80 - 140 | 94 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2 | OK | 2.08 | OK | 80 - 140 | 93 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 14.96 | OK | 1.0 - 1.3 k | 1.26 | OK |
| Feedback B/Ret | 2 | 18 | 25 | OK | 9.18 | OK | 1.0 - 1.3 k | 1.26 | OK | ||
| Feedback C/Ret | 41 | 25 | 28 | OK | 14.18 | OK | 1.0 - 1.3 k | 1.26 | OK | ||
| Feedback D/Ret | 22 | 6 | 35 | OK | 17.38 | OK | 1.0 - 1.3 k | 1.25 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 96,88,82,85,82 | OK | 30 - 40 | 59, 52, 46, 48, 46 | OK | 20 - 30 | 41,30,25,27,25 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 78,78,82,78 | OK | 39, 39, 43, 39 | OK | 20 - 30 | 25,24,29,25 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 76,76,69,76 | OK | 40.5, 40, 33, 40 | OK | 20 - 30 | 22,22,14,21 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 83,80,80,80 | OK | 46, 44, 43, 43 | OK | 20 - 30 | 26,23,23,23 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.061 | OK | 1.0 - 1.1 k | 1.04 | OK | 1.0 - 1.1 k | 1.25 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open, 40 MOhm (C&D) | OK | < 1 nS | >55 MOhm | OK | <1 nS | >55 MOhm | OK | |
| Qo bias/QET Ret | 11 | open, 39 MOhm (C&D) | OK | >55 MOhm | OK | <1 nS | >55 MOhm | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43 MOhm | OK | 6 - 8 nS | 50 MOhm | OK | <1 nS | 53 MOhm | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 43 MOhm | OK | 49 MOhm | OK | <1 nS | 55 MOhm | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.71 | OK | 0.9 - 1.0 V | 1.08 | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.137 | – | 1.2 - 1.5 V | 1.4 | – | — | – | |
| LED 2 | 26 | chassis | 1.127 | – | 1.4 | – | — | – | |||
| 2230 | Fill N2 cold trap full. |
|---|---|
| Open gate valve and condenser line speedivalve. Start condensing mixture through nitrogen cold trap 1. |