2011/12/25

1020 A 'configuration is slow' DAQ error occurred at 08:50.
1023 Found the the above error occurred at 00:57 this morning. All series after that are empty series. The detector control vi was running when taking data, not sure it caused the problem.

Data taken in the last run

# -13V Am-241
2.5G    ./07111224_1521
2.6G    ./07111224_1641
# -8V Am-241
2.8G    ./07111224_1806
2.8G    ./07111224_1927
# -4V Am-241
2.8G    ./07111224_2052
2.8G    ./07111224_2212
# -2V Am-241
2.8G    ./07111224_2337
2.8G    ./07111225_0057

1038 The DAQ error might be due to the run duration array had fewer elements than bias array.
1042 Ramp up tower temperature to 400mK.
1105 Trigger and timing link look fine at 0V test run.
1601 Stop 1K pot, warm up fridge.

Data taken at 400mK

# -13V Am-241
2.5G    ./07111225_1107
# -8V Am-241
2.7G    ./07111225_1232
# +8V Am-241
2.7G    ./07111225_1358
# +13V Am-241
1.6G    ./07111225_1523