2011/12/24
Data taken in the last run
# +8V Am-241 2.6G ./07111223_1628 2.6G ./07111223_1748 2.6G ./07111223_1908 # +10V Am-241 2.8G ./07111223_2033 2.8G ./07111223_2153 2.8G ./07111223_2314 # +13V Am-241 2.8G ./07111224_0039 2.8G ./07111224_0159 2.8G ./07111224_0319 # -3V Am-241 2.7G ./07111224_0444 2.8G ./07111224_0604 2.8G ./07111224_0725
| 1321 | Power off two DCRC's |
|---|---|
| 1349 | Corrected the 0V bias point in the firmware of DCRC32. |
| 1421 | Replaced DCRC30 with DCRC32 and powered up both DCRC33 and DCRC32. The charge amplifier power supply didn't turn on automatically on DCRC32. A temperature spike showed up. |
| 1517 | Created new DAQ config files for DCRC33 and DCRC32. Trigger and timing link look OK in the 0V test run. |
| 1521 | Taking data 07111224_1521, -13V, Am-241. Will take data at -13, -8, -4, -2, +2, +4, +8, +13V. 2 series at each bias. |
| 1523 | Need to use the default phonon offset on DCRC32 instead of DCRC30's. |
| 1538 | Filled N2 cold trap. |