-13V Am-241 Data - Long flashing (10 hrs)

All figures below show order for individual charges: Top row- Q1, Q2; Bottom row- Q3, Q4
Cut applied to the following analyses: Primary charge channel cut
No strict Chi2 cut was applied to spectra - only simple threshold cuts
For brevity on this page, I supplied links to charge, chi2, and charge baseline stdv. vs event time.
-13V Bias
Fig. 1 - Charge (1000 a.u.) vs. Event Time (sec) at -13V Bias
http://www.hep.umn.edu/cdms/cdms_restricted/K100/runs/run022/Voltage_Biases/n13V/20111208/v3_am/q_et.png
Fig. 4 - Chi2 vs. Charge (1000 a.u.) at -13V Bias
Fig. 5 - Charge Spectra (1000 a.u.) at -13V Bias
1000 Filled LN2 cold trap.
1010 DAQ error occurred last night during flashing. It seems the connection to DCRC30 was reset at 03:52.

IGH at 10:12
1017 See pulses when triggering at 40ADC bin. Looks the crystal is recovered.
1021 0V test run 07111208_1020 looks OK on screen.
1023 Start taking 07111208_1023_F0001. Am-241 data at -13V triggering at 40ADC bin. Data/flash/cooldown = 60/0.1/20 min. Tower temperature 61.4mK.
1326 DAQ error occurred at 12:56. Connection to DCRC30 was reset.

1503 Unplugged two boards to let them cool for a while.
1617 Power up both boards.
1619 07111208_1619: flash detector for 6s and then cool down for 20min. Take Ba-133 data at -13V after flashing.
1743 Data not taken because of an error. Flash detector for 0.2min.
1810 Begin LHe transfer (~24%).
1843 End LHe transfer (~92%).
2015 Filled N2 cold trap.
2111 Temperature kept rising and fluctuating since 19:00. G1 and G2 kept rising too. There might be a blockage in the helium cold trap.