2011/11/25

1434 G1=502mbar, G2=609mbar, V6=13.9%. Outlet of N2 cold trap is full of ice. Cryocleaning is running too fast.
1439 IVC pressure 3e-5mbar. Tightened IVC lid again.
1615 Raise dewar. Pump and flush main bath. Evacuate 1 K pot.
Interrupt cryocleaning and pump mixture into dump. Open gate valve and condenser line speedivalve in an effort to remove any remaining contamination in mixing chamber lines.
Fill IVC with 5.4 mbar N2 exchange gas. Start nitrogen transfer.
1845 Nitrogen liquid level meter resistance 1.32 kΩ. Stop nitrogen transfer. Attach ping-pong ball.
2000 Close condenser line speedivalve and gate valve. Recommence cryocleaning of mixture through nitrogen cold trap 2.