UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20111104

2011/11/04

1600 Crystal wire bonded. Pin assignments as follows:
DIB Pin
QET A 2 2
QET A Return 2 1
QET B 1 2
QET B Return 1 1
QET C 2 6
QET C Return 2 5
QET D 1 8
QET D Return 1 7
LED 1 DIB 1 1 9
LED 2 DIB 1 1 10
LED 1 DIB 2 2 9
LED 1 DIB 2 not bonded
Qo-Qi shorted in several places
Qi-Qo bonded to DIB 1 Pin 5
2200 Perform detector level continuity checks.
DIB 1 DIB 2
Signal Measured Measured
QET A - 40.5 Ω
QET B 160 Ω -
QET C - 104 Ω
QET D 371 Ω
LED1 R forward 737 kΩ 725 kΩ
LED1 V forward 1.056 V 1.053 V
LED1 G backward ≈ 5 nS <0.3 nS
LED2 R forward 750 kΩ -
LED2 V forward 1.056 V -
LED2 G backward <0.1 nS -
2215 Shorts found to chassis: suspect Pin 1 DIB 2 is culprit.