| 1111 | 'Not enough memory to complete this operation' error occurred at 21:53 last night. Connection to DCRC30 was also lost. |
| 1122 | Signals on DCRC33 are smaller than that on DCRC30 and sometimes going downward. This maybe because DCRC33 was under bias all the time. |
| 1126 | Flash detector for 10min and the retake -13V Ba data. |
| 1309 | Stop data 07111030_1301 and start data 07111030_1309_F0001. Connection to DCRC30 was lost. Bias -13V with Ba source. take 3 series. |
| 1426 | DAQ error occurred at 07111030_1309_F0014 Event 141898. |
| 1430 | Flash detector for 6s and then take +13V ba data. Tower temperature 58mK. |
| 1558 | +13V data 07111030_1456_F0016 was finished. Raise Tower temperature to 400mK. |
| 1715 | Temperature fluctuates around 400mK with amplitude ~+/- 4mK. Change P=10 and I=120 to P=10 and I=0. |
| 1728 | Temperature fluctuates around 370mK and is noisy. Increase P to 20. |
| 1730 | Temperature rise to ~380mK. Increase P to 100. |
| 1732 | P=200. |
| 1734 | P=400. |
| 1739 | Change heater range from IGH=2mW, LS370=10mA to IGH=20mW and LS370=3.16mA. P=200. |
| 1740 | P=50 |
| 1743 | P=10 |
| 1745 | P=20. |
| 1755 | P=10, I=60 |
| 1808 | Temperature at 400mK is not stable. take some -13V data anyway. 07111030_1808 |
| 1811 | Connection to DCRC 30 was reset. |
| 1814 | start take data again. |
| 1924 | Take +13V data 07111030_1924_F0001 after 6s flashing. |
| 2002 | He bath is dry, stop 1K pot. |
| 2006 | Stop circulation. Pump mixture out from the still side. |
| 2008 | Temperature was kept 400+/-5mK until this moment. Over shoot up to 425mK observed after this point. |
| 2028 | Temperature went up to 440mK at some point of the later data. |