| 0945 | Mixture wasn't circulating in cryocleaning. Increase V6 to 14.1%. MC X12004 = 5.4K. Temperature kept rising, there is not likely a leak. IVC pressure 5.4e-7mbar. |
| 1043 | Pump mixture into dump. Close all valves. |
| 1100 | Close IVC, start 1K pot, open gate valve and condenser valve, start condensing mixture. |
| 1301 | Start to circulate mixture. |
| 1823 | Tower temperature 66.8mK. Power up two DCRC boards. |
| | Start flashing detector. Flashing two LEDs on DIB1, temperature rose to 322mK. Flashing all four LED2, temperature rose to 455mK. Flashing settings: LED currrnt 0.6mA, 100us ON, 5000us OFF, both LEDs enabled. |
| 1841 | Tower temperature dropped to 361mK at 18:36. Found LEDs on DCRC30(DIB1) stopped working. It was likely the board was reset. |
| 1847 | Filled N2 cold trap. |