2011/10/17

1459 Dewar pressure 0.25mbar. Start pumping the dewar.
1515 Dewar pressure 7.4E-3 mbar.
1700 Dewar pressure 1.6e-3 mbar
1900 Tower assembly complete. Conduct tower level continuity checks.
Detector G102 DIB 1
side coax ZB1-4-8 date/who: 2011-10-17 HC & AC
face 2
+ve lead -ve lead signal measured
3 4 QET B 126.9 Ω
9 chassis LED1 Rforward 780 kΩ
10 chassis LED2 Rforward 755 kΩ
9 chassis LED1 Vforward 1.102 V
10 chassis LED2 Vforward 1.101 V
chassis 9 LED1 Rreverse <0.3 nS
chassis 10 LED2 Rreverse <0.5 nS
3 11 Qo Rbias 25.0 nS
12 11 Qo C coupling <0.1 nS
12 13 Qo Rfeedback 25.5 nS
3 14 Qi Rbias 26.0 nS
15 14 Qi C coupling <0.1 nS
15 16 Qi Rfeedback 25.0 nS
Detector G102 DIB 2
side coax ZB1-4-6 date/who: 2011-10-17 HC & AC
face 5
+ve lead -ve lead signal measured
5 6 QET C 337 Ω
6 8 QET C-D 362 Ω
7 8 QET D 1.34 kΩ
9 chassis LED1 Rforward 774 kΩ
10 chassis LED2 Rforward 752 kΩ
9 chassis LED1 Vforward 1.108 V
10 chassis LED2 Vforward 1.106 V
chassis 9 LED1 Rreverse <0.2 nS
chassis 10 LED2 Rreverse <1.0 nS
5 11 Qo Rbias 25.5 nS
12 11 Qo C coupling <0.1 nS
12 13 Qo Rfeedback 26.0 nS
5 14 Qi Rbias 25.0 nS
15 14 Qi C coupling <0.1 nS
15 16 Qi Rfeedback 25.5 nS