1745 Fill N2 cold trap full.
Connect He leak checker. Background leak rate very small, falls off-scale. Perhaps checker might require recalibration. It is pumping on a very small volume at the moment though.
1755 IVC pressure 5.4E-7 mbar. Isolate turbo pump and open valve to leak checker. Background He leak rate jumps to 1E-7 mbar l/s. Helium background is a little high, but this seems to indicate that the leak checker is working.
1852 No gross leak found from circulation line to IVC at 9.8K and He BG 1.6E-7cc/s.
1920 Start 1 K pot. Isolate leak checker: readings are too low as 1 K pot acts like cryopump, and hence leak checker's pump is less efficient. If mixture doesn't condense, there is a leak.
Start condensing mixture.
2145 Mixture condensed a while ago, and mixing chamber temperature has plateaued at 2.2 K.
Start circulation.
2200 Mixing chamber temperature has dropped below 1.2 K. In absence of tower thermometer, switch to still thermometer to get an idea of the base temperature.