| 1122 | Found MC X12004 started to drop from 7.1K at 23:45 last night and reached and stayed at 4.3K from 01:06 this morning. IVC pressure 4.6e-6mbar. There might still be a small leak. I'll start condensing mixture and see. |
| 1135 | Filled N2 cold trap. |
| 1141 | Start 1K pot, start condensing mixture. 1K pot temperature 1.622K. |
| 1256 | 1K pot 1.818K. G1 and G2 kept rising. There might be a blockage in the circulation line. Pump mixture out from the still line. G1 dropped. The circulation line is still open. |
| 1300 | MC temperature rises while pumping the mixture. G1 and G2 also rise. P1 stays at 0.55mbar. |
| 1312 | Pump IVC, IVC pressure 5.4e-5mbar. |
| 1315 | G1 is dropping. There seems too much Helium left in the IVC and it thermally shorts the fridge below 2.2K. |
| 1341 | IVC pressure rose to 5.4e-6mbar from 3e-6mbar. MC X12004 rose from 3.19K. |
| 1342 | MC X12004 3.45K. IVC P = 7.4E-6 mbar. |
| 1343 | IVC P = 8.6e-6 mbar. |
| 1351 | MC temperature and IVC pressure both went down. |
| 1526 | IVC pressure 4.6e-6mbar. MC X12004 3.49K. 1K pot 1.708K. |
| 1733 | IVC temperature fluctuates with temperature. There must be fairly amount of He in the IVC from a leak. Now stop 1K pot and pump out mixture. |
| 1738 | Stopped 1K pot, added 2mW to MC, pumping mixture into dump. |