| 1150 | DAQ returned memory error again at start of first file of series 07110905_0531. Stopped DAQ. |
Data taken in the last run with reasonable size, 12 series total.
941M ./07110904_1908
974M ./07110904_1959
952M ./07110904_2050
969M ./07110904_2141
1.2G ./07110904_2236
1.2G ./07110904_2327
1.2G ./07110905_0017
1.2G ./07110905_0108
1.3G ./07110905_0204
1.2G ./07110905_0255
1.3G ./07110905_0345
1.2G ./07110905_0436
| | Tower temperature increased sharply at 0830, reaching 790 mK and rising at a decent pace. Temperature of 1 K pot at 4.5 K. Pressures in G1 & G2 relatively low at just under 40 mbar. |
| | The above would indicate that the 1 K pot pump has failed, and the mixture is not circulating (perhaps due to the failure of the He-3 pump). These suspicions are vindicated upon checking the pumps - both pumps are off! It doesn't look like a general power failure as DAQ computer was not rebooted. |
| | Upon pushing the mains switch, the 'DC FAIL' light illuminates. Resting the power unit seems to do the trick - both pumps whirl into action. |
| | Tower temperature begins to drop. Keep eye on 1 K pot temperature and pressures in G1 & G2. |
| 1235 | Fill N2 cold trap full. |
| 1330 | A tower temperature of 55.5 mK reached. Tower temperature still falling. |
| 1521 | A connection refused error occurred when trying to start the DAQ to take data. Found port 5001 of DCRC30 was not released. |
| 1525 | After releasing port 5001 of DCRC30. The DAQ worked fine. |
| 1547 | Start data 07110905_1547_F0001. -6V on Q1-3. Continuation of last run. 1st series of -6V data. |