| 0853 | A DAQ error occurred when quitting the DCRCv3MultiDAQ.vi at ~ 01:50. The last series of data might be fine. |
| 0910 | DCRC30 trigger seem to lose response. The two charge amplifier LEDs are off. Cycle the power of it. |
| 0915 | DCRC30 came back online. Tower temperature 54.7mK. |
| 0918 | Timing link is on. Flash the detector for 10 min since the detector has been under -5V bias after the DAQ error. Take -7V data after 80min cooldown. |
| 0932 | Filled N2 cold trap. |
| 1157 | DCRC30 lost signal. This may happen at 10:54 where a temperature spike showed up. Stop the DAQ. Found the charge amp LEDs are off. Cycle the power. |
Data taken
1.3G ./07110902_1053
522M ./07110902_1144
The data could not be processed. Just deleted the files.
| 1217 | DCRC30 lost charge amp power again when trying take data. Was it because the bias was too large or POE problem? |
| 1753 | Flash 10 min and then take +1, +2, +4, +6V data. 4 series for each bias. Trigger on Q1-3 at 32 ADC bins. Data/flash/cooldown = 30/0.1/20 min. |