2011/08/28
| 0130 | Piping cleaned and reassembled. |
|---|---|
| Recommence evacuating main bath. Pressure in P2 decreases below 1 mbar, indicating that there is no remaining liquid nitrogen. | |
| 0150 | Fill IVC with 1.4 mbar He-4 exchange gas. |
| Conduct electrical check at 77 K - no problems. | |
| 0205 | Start first LHe transfer. |
| 0215 | Fill N2 cold trap full. |
| 0325 | Pump out He-4 exchange gas at mixing chamber temperature of 20 K. |
| 1650 | Fill N2 cold trap full. |
| Close gate valve and condenser line speedivalve. Close all valves on IGH. Start cryocleaning mixture through cold trap 2; open valve 6 to 13.2%. | |
| 1710 | Mixture has stagnated. Open valve 6 to 14.0%. |
| 1735 | Conduct electrical check at 5.5 K. |
| Detector: | G102 DIB1 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-4-6 | 300 K | 80K | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2011-08-26 | Date | 2011-08-28 | Date | 2011-08-28 | ||||
| SQUET: | ZC23 | Who | HC&JZ | Who | HC&JZ | Who | HC&JZ | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.7 | OK | 80 - 140 | 93 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2 | OK | 80 - 140 | 94 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.8 | OK | 80 - 140 | 101 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.7 | OK | 80 - 140 | 92 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18 | OK | 9 - 13 k | 7.6 | OK | 1.0 - 1.3 k | 1.28 | OK |
| Feedback B/Ret | 2 | 18 | 20 | OK | 8.6 | OK | 1.0 - 1.3 k | 1.28 | OK | ||
| Feedback C/Ret | 41 | 25 | 19 | OK | 8 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7.4 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 85,106,85,86,86 | OK | 30 - 40 | 50,70,48,49,49 | OK | 20 - 30 | 32,56,31,32,32 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 86,85,87,86 | OK | 49,48,50,49 | OK | 20 - 30 | 31,30,32,31 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 88,85,74,84 | OK | 52,49,37,47 | OK | 20 - 30 | 34,32,20,30 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 84,84,84,84 | OK | 49,48,48,48 | OK | 20 - 30 | 32,31,31,31 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.043 | OK | 1.0 - 1.1 k | 1.27 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 38 MOhm | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 38 MOhm | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | OL | OK | 6 - 8 nS | OL | OK | <1 nS | OK | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | OL | OK | OL | OK | <1 nS | OK | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.87 | OK | 0.9 - 1.0 V | 1.141 | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.259 | – | 1.2 - 1.5 V | 1.483 | – | — | – | |
| LED 2 | 26 | chassis | 1.239 | – | 1.466 | – | — | – | |||
| Detector: | G102 DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-4-8 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2011-08-26 | Date | 2011-08-28 | Date | 2011-08-28 | ||||
| SQUET: | ZC70 | Who | HC&JZ | Who | HC&JZ | Who | HC&JZ | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8 | OK | 1 - 2 k | 2 | OK | 80 - 140 | 96 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.8 | OK | 80 - 140 | 102 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.9 | OK | 80 - 140 | 93 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2 | OK | 2.2 | OK | 80 - 140 | 92 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 15 | OK | 1.0 - 1.3 k | 1.26 | OK |
| Feedback B/Ret | 2 | 18 | 26 | OK | 10 | OK | 1.0 - 1.3 k | 1.26 | OK | ||
| Feedback C/Ret | 41 | 25 | 28 | OK | 15 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback D/Ret | 22 | 6 | 35 | OK | 18 | OK | 1.0 - 1.3 k | 1.25 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 80,100,79,80,79 | OK | 30 - 40 | 48,68,45,46,46 | OK | 20 - 30 | 25,45,22,23,23 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 76,76,77,76 | OK | 40,40,41,40 | OK | 20 - 30 | 24,23,25,24 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 80,77,69,76 | OK | 47,44,36,43 | OK | 20 - 30 | 26,23,15,22 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 80,80,80,80 | OK | 47,47,46,46 | OK | 20 - 30 | 25,24,24,24 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06 | OK | 1.0 - 1.1 k | 1.04 | OK | 1.0 - 1.1 k | 1.26 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 38 MOhm | OK | < 1 nS | OK | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 39 MOhm | OK | OK | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | OL | OK | 6 - 8 nS | OL | OK | <1 nS | OK | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | OL | OK | OL | OK | <1 nS | OK | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | open | OK | — | OK | |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.235 | – | 1.2 - 1.5 V | 1.466 | – | — | – | |
| LED 2 | 26 | chassis | 1.218 | – | 1.455 | – | — | – | |||
| 1740 | Stop cryocleaning mixture, and pump it into dump. |
|---|---|
| Start 1 K pot. | |
| 1800 | Start condensing mixture through cold trap 1. |
| 1945 | Mixture condensed. Start circulation. |
| IVC pressure 8.6E-7 mbar. Isolate IVC. | |
| 2010 | Valve 6 fully open, mixture out of dump and circulating. |
| 2015 | Close needle valve a little to maintain 1 K pot temperature between 1.5 and 1.6 K. |
| 2035 | Flash and bake detector overnight. |