2011/08/25
| 1630 | Tower assembled and tower level continuity checks conducted. |
|---|
| Detector | G102 | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-4-6 | date/who: | 2011-08-25 | JZ |
| face | 2 | |||
| +ve lead | -ve lead | signal | measured | |
| 3 | 4 | QET B | 136.7 Ω | |
| 9 | chassis | LED1 Rforward | 800 kΩ | |
| 10 | chassis | LED2 Rforward | 800 kΩ | |
| 9 | chassis | LED1 Vforward | 1.110 V | |
| 10 | chassis | LED2 Vforward | 1.109 V | |
| chassis | 9 | LED1 Rreverse | ||
| chassis | 10 | LED2 Rreverse | ||
| 3 | 11 | Qo Rbias | 25.5 nS | |
| 12 | 11 | Qo C coupling | <0.1 nS | |
| 12 | 13 | Qo Rfeedback | 25.8 nS | |
| 3 | 14 | Qi Rbias | 25.1 nS | |
| 15 | 14 | Qi C coupling | <0.1 nS | |
| 15 | 16 | Qi Rfeedback | 25.6 nS | |
| Detector | G102 | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-4-8 | date/who: | 2011-08-25 | JZ |
| face | 5 | |||
| +ve lead | -ve lead | signal | measured | |
| 5 | 6 | QET C | 343 Ω | |
| 6 | 8 | QET C-D | 354 Ω | |
| 7 | 8 | QET D | 1.46 kΩ | |
| 9 | chassis | LED1 Rforward | 800 kΩ | |
| 10 | chassis | LED2 Rforward | 800 kΩ | |
| 9 | chassis | LED1 Vforward | 1.105 V | |
| 10 | chassis | LED2 Vforward | 1.103 V | |
| chassis | 9 | LED1 Rreverse | <0.3 nS | |
| chassis | 10 | LED2 Rreverse | <1.2 nS | |
| 1 | 11 | Qo Rbias | 25.3 nS | |
| 12 | 11 | Qo C coupling | <0.1 nS | |
| 12 | 13 | Qo Rfeedback | 25.5 nS | |
| 1 | 14 | Qi Rbias | 25.8 nS | |
| 15 | 14 | Qi C coupling | <0.1 nS | |
| 15 | 16 | Qi Rfeedback | 25.2 nS | |
| 1800 | Commence installation in K100. |
|---|