| 0919 | Stop flashing detector. |
| 1026 | Tower temperature plateaued at 103mK. Temperature dropped quickly below 80mK after resetting the DCRC boards. |
| 1044 | Filled N2 cold trap. |
| 1736 | Zap squids of ZC70 on DCRC30. |
| 1743 | Zap squids on ZC23, DCRC33. |
| 2017 | Start to take data 07110718_2017_F0001. DIB1 -2.5V, DIB2 2.5V. Trigger on charge. Trigger rate is very low. |
| 2053 | Found that the DAQ trigger rate was too low was because the DCRC trigger rate was too high. Set trigger threshold to 64 ADC bins solved the problem. |
| 2055 | Start taking data 07110718_2055_F0001. DIB1 -2.5V ZIP1, DIB2 2.5V ZIP2. Trigger on all 4 charge channels at 64 ADC bin. Ba-133 source. Data/flash/cooldown=120min/6s/20min. |
| 2106 | Connection to DCRC30 was reset and an DAQ error occurred. |