| 1055 | MC temperature 77.2K. IVC pressure:1.2mbar. |
| 1115 | Start to pump out exchange gas. |
| 1123 | IVC He background: 2.3e-9cc/s with pumping station connected. |
| 1328 | Cryo-cleaning was running too fast. Ice was condensed on the cold trap outlet. Adjusted V6 to 12.5%. |
| 1458 | Cryo-cleaning seemed stopped. Increase V6 to 13.9%. |
| 1537 | Pump mixture into dump. |
| 1603 | Isolate N2 cold trap 2 at G1=0.1mbar. G2=610mbar. Close V6. |
| 1610 | Isolate pumping station at IVC P=3e-6mbar. |
| 1624 | Filled N2 cold trap. IVC HE background: 9.7e-9cc/s. |
| 1631 | Throughput test: 50mbar/min. NO gross leak at BG 9.7e-9cc/s. Pump mixture into dump. |
| 1645 | Start to blow out N2 |
| 1709 | Electrical checks are OK at 77K. |
| 1800 | Liquid N2 is out. G3 starts to drop. |
| 1811 | Stop pressurizing the bath. IVC He signal: 1.1e-8cc/s. |
| 1825 | Start to pump the main bath. |
| 1838 | Put in 0.16mbar He-4 exchange gas in IVC. |
| 1839 | Fill the bath with He gas. |
| 1844 | Evacuated and isolated 1K pot. |
| 1908 | Start initial He transfer. |
| 1957 | Start to pump out He-4 exchange gas at 25K, IVC pressure: 0.025 mbar. Temperature dropping rate was 1.25K/min. |
| 2009 | Temperature reaches the lowest point 15.3K and starts to rise. |
| 2049 | IVC He: 2.7e-6cc/s with pumping station connected. |
| 2059 | Finished He transfer at 91.2%. IVC He level 2.5e-6cc/s. |
| 2110 | IVC He level 1.96e-6cc/s. |