2011/07/14

1900 Finish assembling tower. No shorts to chassis. Perform tower level continuity check.
Detector G16K DIB 1
side coax ZB1-1-9 date/who: 2011-07-14 HC
face 2
+lead - lead signal typical measured
1 2 QET A 100 to 200 128
2 4 QETA-B 100 to 200 245
3 4 QET B 100 to 200 243
4 6 QETB-C 100 to 200 268
5 6 QET C 100 to 200 207
6 8 QETC-D 100 to 200 265
7 8 QET D 100 to 200 200
2 8 QETA-D 100 to 200 231
chassis 9 LED1 Rforward 8M to 9M 787k
chassis 10 LED2 Rforward 8M to 9M 783k
chassis 9 LED1 Vforward 1.0 V to 1.1 V 1.105 V
chassis 10 LED2 Vforward 1.0 V to 1.1 V 1.103 V
9 chassis LED1 Rreverse <1 nS <0.2 nS
10 chassis LED2 Rreverse <1 nS <0.5 nS
1 11 Qo Rbias 24 nS to 26 nS 25 nS
12 11 Qo C coupling <1 nS <0.1 nS
12 13 Qo Rfeedback 24 nS to 26 nS 26 nS
1 14 Qi Rbias 24 nS to 26 nS 26 nS
15 14 Qi C coupling <1 nS <0.2 nS
15 16 Qi Rfeedback 24 nS to 26 nS 26 nS
Detector G16K DIB 2
side coax ZB1-1-6 date/who: 2011-07-14 HC
face 5
+lead - lead signal typical measured
1 2 QET A 100 to 200 126
2 4 QETA-B 100 to 200 230
3 4 QET B 100 to 200 225
4 6 QETB-C 100 to 200 268
5 6 QET C 100 to 200 208
6 8 QETC-D 100 to 200 249
7 8 QET D 100 to 200 201
2 8 QETA-D 100 to 200 207
chassis 9 LED1 Rforward 8M to 9M 778k
chassis 10 LED2 Rforward 8M to 9M 784k
chassis 9 LED1 Vforward 1.0 V to 1.1 V 1.104 V
chassis 10 LED2 Vforward 1.0 V to 1.1 V 1.104 V
9 chassis LED1 Rreverse <1 nS <0.2 nS
10 chassis LED2 Rreverse <1 nS <0.2 nS
1 11 Qo Rbias 24 nS to 26 nS 26 nS
12 11 Qo C coupling <1 nS <0.1 nS
12 13 Qo Rfeedback 24 nS to 26 nS 26 nS
1 14 Qi Rbias 24 nS to 26 nS 26 nS
15 14 Qi C coupling <1 nS <0.3 nS
15 16 Qi Rfeedback 24 nS to 26 nS 26 nS