2011/07/13
| 1800 | Finish assembling tower. |
|---|---|
| Perform tower level continuity check. |
| Detector | G16K | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2011-07-13 | HC & AK |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 126 |
| 2 | 4 | QETA-B | 100 to 200 | 245 |
| 3 | 4 | QET B | 100 to 200 | 241 |
| 4 | 6 | QETB-C | 100 to 200 | 266 |
| 5 | 6 | QET C | 100 to 200 | 204 |
| 6 | 8 | QETC-D | 100 to 200 | 262 |
| 7 | 8 | QET D | 100 to 200 | 197 |
| 2 | 8 | QETA-D | 100 to 200 | 230 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 789k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 774k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.103 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.102 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.2 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.5 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 25.5 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 25.5 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 25.5 nS |
| Detector | G16K | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-1-6 | date/who: | 2011-07-13 | HC & AK |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 128 |
| 2 | 4 | QETA-B | 100 to 200 | 233 |
| 3 | 4 | QET B | 100 to 200 | 249 |
| 4 | 6 | QETB-C | 100 to 200 | 272 |
| 5 | 6 | QET C | 100 to 200 | 210 |
| 6 | 8 | QETC-D | 100 to 200 | 254 |
| 7 | 8 | QET D | 100 to 200 | 204 |
| 2 | 8 | QETA-D | 100 to 200 | 209 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 780k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 786k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.105 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.105 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.2 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.2 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.3 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 26 nS |