2011/06/29
| 1730 | Assemble tower and perform tower level continuity checks. |
|---|
| Detector | G19N | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2011-06-29 | HC & AK |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 129.8 |
| 2 | 4 | QETA-B | 100 to 200 | 249.6 |
| 3 | 4 | QET B | 100 to 200 | 248.3 |
| 4 | 6 | QETB-C | 100 to 200 | 275.0 |
| 5 | 6 | QET C | 100 to 200 | 212.0 |
| 6 | 8 | QETC-D | 100 to 200 | 267.3 |
| 7 | 8 | QET D | 100 to 200 | 201.8 |
| 2 | 8 | QETA-D | 100 to 200 | 233.0 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 801k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 775k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.106 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.106 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <1.0 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.4 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.6 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 25 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.5 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 25 nS |
| Detector | G19N | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-1-6 | date/who: | 2011-04-23 | HC & AK |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 131.5 |
| 2 | 4 | QETA-B | 100 to 200 | 259.2 |
| 3 | 4 | QET B | 100 to 200 | 256.2 |
| 4 | 6 | QETB-C | 100 to 200 | 275.4 |
| 5 | 6 | QET C | 100 to 200 | 210.5 |
| 6 | 8 | QETC-D | 100 to 200 | 274.6 |
| 7 | 8 | QET D | 100 to 200 | 209.4 |
| 2 | 8 | QETA-D | 100 to 200 | 241.7 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 796k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 800k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.109 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.108 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.7 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.5 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <2.0 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.8 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 26 nS |
| 1805 | Start tower installation procedure. |
|---|---|
| 2100 | Tower thermometer has lower than expected resistance. Looks like a fault. Will have to remove temperature stages for a closer inspection. |
| 2300 | Severed wire at connector on tower side. Can be soldered back after removing sealant. |