2011/06/14
| 1626 | -6V bias data is finished. Stop the DAQ after flashing to transfer liquid helium. |
|---|
-8V data
07110613_2348 7 07110614_0038 7 07110614_0128 7 07110614_0218 7 07110614_0309 7 07110614_0359 7 07110614_0449 7 07110614_0540 7 07110614_0630 7 07110614_0721 7
-6V data
07110614_0813 10 07110614_0904 8 07110614_0954 8 07110614_1044 8 07110614_1134 8 07110614_1225 8 07110614_1315 8 07110614_1405 8 07110614_1456 8 07110614_1546 8
| 1637 | Helium transfer started with Helium levels at 8.8%. |
|---|---|
| 1720 | Helium transfer end with Helium levels at 92.1%. |
| 1748 | Filled LN2 Cold Trap. |
| 1802 | Start taking data 07110614_1802_F0001. True -2V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 62.3mK. This is one bias in the bias sweep. |