UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20110613_ba

Table of Contents

2011/06/13

Data Collected in the Last Run

07110605_1721

Information about run

Am-241 & Ba-133 

-2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins.  
Trace length 3.2768 ms (8192 samples), lookback time 400 μs; ozip format. 

Additionally, no flashing was done during this six hour run.

Cuts Used

QiCut=inrange(QIOFvolts(401,0)./(QIOFvolts(401,0)+QOOFvolts(401,0),.8,1.2&inrange(QIOFvolts(402,0)./(QIOFvolts(402,0)+QOOFvolts(402,0),.8,1.2)

chiSqCut = QSOFchisq(401,0)<(2.5e8*(QIOFvolts(401,0)+QOOFvolts(401,0)).^2+18000)&QSOFchisq(402,0)<(2.5e8*(QIOFvolts(402,0)+QOOFvolts(402,0)).^2+18000)

Data Plots

Figure 1: Qsum from data series 07110605_1721.
Figure 2: Phonon Response. Outer channels not included.
Figure 4: Phonon Energy. Note it has not been normalized. Figure 5: Qzpart as a function of time.
Figure 6: Yield as a function of time.
Figure 7: Percent of Yield above .3e-4. Figure 8: Percent of events passing Chi Squared Cut.

Findings

It seems that our detectors are maintaining yield much longer than those at UC-Berkeley as it remained constant for approximately 3 hours. It is important to note that as the percent of yield stays constant for the first three hours, the percent of events passing the chi squared cut decreases immediately.