| 1605 | Fill N2 cold trap full. |
| 1625 | Stop flashing detector. |
| 1825 | Start taking data 07110611_1825_F0001. True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 2h/6s/30min. Tower temperature 68.3mK. Trigger baseline/sample = 64/2. |
| 1857 | Stopped the run at 07110611_1825_F0007 to transfer liquid helium. Set charge biases to 0 |
| 1910 | Begun LHe transfer (29.4%). |
| 2019 | Start taking data 07110611_2019_F0001. True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 67.1mK. Trigger baseline/sample = 64/16. Timing link wasn't set for the first 1300 events. |
| 2149 | Stop the run after flashing. Series 07110611_2019 and 07110611_2109 were taken. |