2011/06/11

1605 Fill N2 cold trap full.
1625 Stop flashing detector.
1825 Start taking data 07110611_1825_F0001. True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 2h/6s/30min. Tower temperature 68.3mK. Trigger baseline/sample = 64/2.
1857 Stopped the run at 07110611_1825_F0007 to transfer liquid helium. Set charge biases to 0
1910 Begun LHe transfer (29.4%).
2019 Start taking data 07110611_2019_F0001. True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 67.1mK. Trigger baseline/sample = 64/16. Timing link wasn't set for the first 1300 events.
2149 Stop the run after flashing. Series 07110611_2019 and 07110611_2109 were taken.

Data taken in the last run:

07110611_2019 7
07110611_2109 8
Adjusted Q1 trigger threshold to have all the channels triggering at ~22Hz.
2210 Start taking data 07110611_2210_F0001. Trigger testing run. True -4V on Q1-4, Trigger on Q1-4 at 0x(24, 40, 40, 40)ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 68.6mK. Trigger baseline/sample = 64/16.

Data

Check charge baseline std