2011/06/09
| 0115 | Fill N2 cold trap full. |
|---|---|
| 1046 | MC X12004 79.28K, IVC P=2.5mbar. Start pumping out N2 exchange gas. |
| 1054 | Pump IVC with leak checker down to 7e-2mbar and then open the turbo pump. He background: 2.5e-9cc/s with pumping station connected. |
| 1226 | Open V6 13.8%→14%, there wasn't gas flowing through. G2=460mbar, G1=452.3mbar. |
| 1350 | Interrupt cryocleaning. Pump mixture into dump in preparation for throughput test. |
| 1424 | Perform electrical check at 77 K. |
| Detector: | G42 DIB1 | BH: | Stripline: | 19 | ||||
| SC: | ZB1-1-9 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2011-06-07 | Date | 2011-06-09 | |||
| SQUET: | ZC23 | Who | HC & JZ | Who | HC, AK & RWR | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.6 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.7 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.6 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18 | OK | 9 - 13 k | 7.3 | OK |
| Feedback B/Ret | 2 | 18 | 20 | OK | 8.2 | OK | ||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.6 | OK | ||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7.1 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 84,86,84,85,86 | OK | 30 - 40 | 45,58,45,46,46 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 85,85,88,85 | OK | 46,45,46,45 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 85,85,74,84 | OK | 45,45,34,44 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 84,83,83,84 | OK | 45,45,44,45 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.044 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open (A,C,D) 38MOhm (B) | OK | < 1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | open (A,C,D) 39MOhm (B) | OK | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 44 MOhm, 44 MOhm | OK | 6 - 8 nS | 50MOhm, 50MOhm | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 44 MOhm, 44 MOhm | OK | 50MOhm, 50MOhm | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.724 | OK | 0.9 - 1.0 V | 1.091 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.14 | – | 1.2 - 1.5 V | 1.416 | – |
| LED 2 | 26 | chassis | 1.129 | – | 1.406 | – | ||
| Detector: | G42 DIB2 | BH: | Stripline: | 20 | ||||
| SC: | ZB1-1-6 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2011-06-07 | Date | 2011-06-09 | |||
| SQUET: | ZC70 | Who | HC & JZ | Who | HC, AK & RWR | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8 | OK | 1 - 2 k | 2 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.7 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.8 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2 | OK | 2.1 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 15 | OK |
| Feedback B/Ret | 2 | 18 | 25 | OK | 9.2 | OK | ||
| Feedback C/Ret | 41 | 25 | 28 | OK | 14 | OK | ||
| Feedback D/Ret | 22 | 6 | 35 | OK | 17 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 79,81,79,89,80 | OK | 30 - 40 | 44,50,43,44,43 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 76,75,79,76 | OK | 38,37,38,38 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 77,76,69,76 | OK | 41,41,33,40 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 80,80,79,80 | OK | 44,44,43,44 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06 | OK | 1.0 - 1.1 k | 1.04 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open (A&B), 38 MOhm (C&D) | OK | < 1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | open (A&B), 38 MOhm (C&D) | OK | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43 MOhm, 43 MOhm | OK | 6 - 8 nS | 49 MOhm, 49 MOhm | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 44 MOhm, 43 MOhm | OK | 48 MOhm, 49 MOhm | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | 1.085 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.135 | – | 1.2 - 1.5 V | 1.413 | – |
| LED 2 | 26 | chassis | 1.126 | – | 1.405 | – | ||
| 1500 | Conduct throughput test: 30 mbar/min. Fine. |
|---|---|
| Helium leak rate 3.5E-9 mbar l/s: no leak from mixing chamber lines. | |
| 1515 | Start blowing out nitrogen. Helium leak level 3.5E-9 mbar l/s. |
| 1530 | Fill N2 cold trap full. |
| 1610 | Nitrogen is out of main bath. Helium leak level 9.4E-9 mbar l/s, no spike: looks good. |
| Evacuate main bath. | |
| 1640 | Fill IVC with 16 mbar He-4 exchange gas. |
| 1700 | Start first LHe transfer. |
| 1800 | Pump out exchange gas. |
| 1940 | Stop LHe transfer. Liquid helium level 92.8%. |
| Start 1 K pot. | |
| 1955 | Start condensing mixture through nitrogen cold trap 1. |
| 2230 | Mixture has mostly condensed. Mixing chamber temperature is a little high at 4.8 K. Still is cold enough to start circulation. Start circulation. |