1210 Stop pumping out nitrogen cold traps.
Start pumping out mixing chamber lines. Switch on He-3 pump and open valve 9 to dump.
1225 Fill N2 cold trap full and place nitrogen cold traps back in their bath.
1430 Tower assembled: 1 hour on clean bench. Stored in purge cabinet.

Tower Level Continuity Check

1630 Tower removed from purge cabinet. Adapter plate fitted. Perform tower level continuity checks.
Detector G102 DIB 1
side coax ZB1-1-9 date/who: 2011-06-07 HC, MF & AK
face 2
+ve lead -ve lead signal measured
3 4 QET B 129 Ω
9 chassis LED1 Rforward 780 kΩ
10 chassis LED2 Rforward 760 kΩ
chassis 9 LED1 Rreverse <0.6 nS
chassis 10 LED2 Rreverse <2 nS
1 11 Qo Rbias 26 nS
12 11 Qo C coupling <0.3 nS
12 13 Qo Rfeedback 26 nS
1 14 Qi Rbias 26 nS
15 14 Qi C coupling <0.5 nS
15 16 Qi Rfeedback 26 nS
Detector G102 DIB 2
side coax ZB1-1-6 date/who: 2011-06-07 HC, MF & AK
face 5
+ve lead -ve lead signal measured
5 6 QET C 347 Ω
7 8 QET D 1.45 kΩ
9 chassis LED1 Rforward 770 kΩ
10 chassis LED2 Rforward 760 kΩ
chassis 9 LED1 Rreverse <1 nS
chassis 10 LED2 Rreverse <2 nS
1 11 Qo Rbias 26 nS
12 11 Qo C coupling <0.1 nS
12 13 Qo Rfeedback 26 nS
1 14 Qi Rbias 26 nS
15 14 Qi C coupling <0.3 nS
15 16 Qi Rfeedback 26 nS
1700 Tower bagged and taken to mini-clean room.
1830 Begin installation in K100.

Electrical Check at Room Temperature

Detector: G42 DIB1 BH:
SC: ZB1-1-9 300 K
Face: 2 Date 2011-06-07
SQUET: ZC23 Who HC & JZ
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.4 OK
SQUID bias B/Ret 19 34 5 OK
SQUID bias C/Ret 8 24 4.6 OK
SQUID bias D/Ret 23 39 4.3 OK
Feedback A/Ret 37 21 15 - 25 k 18 OK
Feedback B/Ret 2 18 20 OK
Feedback C/Ret 41 25 19 OK
Feedback D/Ret 22 6 18 OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 84,86,84,85,86 OK
QET B/Ret 35 29, 33, 47, 48 85,85,88,85 OK
QET C/Ret 7 5, 9, 10, 28 85,85,74,84 OK
QET D/Ret 38 36, 40, 43, 44 84,83,83,84 OK
FET heater +/- 45 46 1.0 - 1.1 k 1.068 OK
Qi bias/QET Ret 15 24 - 26 nS open (A,C,D) 38MOhm (B) OK
Qo bias/QET Ret 11 open (A,C,D) 39MOhm (B) OK
Qi feedback/Drain/Source 50 16, 31 15 - 17 nS 44 MOhm, 44 MOhm OK
Qo feedback/Drain/Source 30 12, 27 44 MOhm, 44 MOhm OK
Qi drain 16 chassis < 0.1 nS – OK
Qo drain 12 chassis – OK
Qi source 31 chassis – OK
Qo source 27 chassis – OK
FET Tsense +/FET heater- 49 46 0.6 - 0.7 V 0.724 OK
LED 1 42 chassis 1.0 - 1.1 V 1.14 –
LED 2 26 chassis 1.129 –
Detector: G42 DIB2 BH:
SC: ZB1-1-6 300 K
Face: 5 Date 2011-06-07
SQUET: ZC70 Who HC & JZ
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 4.8 OK
SQUID bias B/Ret 19 34 4.4 OK
SQUID bias C/Ret 8 24 4.6 OK
SQUID bias D/Ret 23 39 5.2 OK
Feedback A/Ret 37 21 15 - 25 k 29 OK
Feedback B/Ret 2 18 25 OK
Feedback C/Ret 41 25 28 OK
Feedback D/Ret 22 6 35 OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 79,81,79,89,80 OK
QET B/Ret 35 29, 33, 47, 48 76,75,79,76 OK
QET C/Ret 7 5, 9, 10, 28 77,76,69,76 OK
QET D/Ret 38 36, 40, 43, 44 80,80,79,80 OK
FET heater +/- 45 46 1.0 - 1.1 k 1.06 OK
Qi bias/QET Ret 15 24 - 26 nS open (A&B), 38 MOhm (C&D) OK
Qo bias/QET Ret 11 open (A&B), 38 MOhm (C&D) OK
Qi feedback/Drain/Source 50 16, 31 15 - 17 nS 43 MOhm, 43 MOhm OK
Qo feedback/Drain/Source 30 12, 27 44 MOhm, 43 MOhm OK
Qi drain 16 chassis < 0.1 nS – OK
Qo drain 12 chassis – OK
Qi source 31 chassis – OK
Qo source 27 chassis – OK
FET Tsense +/FET heater- 49 46 0.6 - 0.7 V open OK
LED 1 42 chassis 1.0 - 1.1 V 1.135 –
LED 2 26 chassis 1.126 –