2011/06/04
| 1420 | Stop taking data with neutron source. Files collected (Cf-252, shielded): 07110603_2047 07110603_2139 07110603_2231 07110603_2323 07110604_0015 07110604_0107 07110604_0157 07110604_0247 07110604_0336 07110604_0427 07110604_0518 07110604_0608 07110604_0658 07110604_0749 07110604_0840 07110604_0931 07110604_1022 07110604_1113 07110604_1203 07110604_1253 07110604_1343 |
|---|---|
| 1455 | Begun LHe transfer (~11%). |
| 1530 | End LHe transfer (70.4%). |
| 1710 | Start 60 min/6 s/15 min Ba-133 data series, triggering on phonon channels, +2.5 V DIB1, -2.5 V DIB2, 0.4096 ms trace length, 100 μs trigger delay: data series 07110604_1708 onwards. |
| 2200 | Stop Ba-133 data series. Files collected: 07110604_1708 07110604_1823 07110604_1938 07110604_2053 |
| 2215 | Fill N2 cold trap full. |
| 2235 | Start 30 min/6 s/15 min data series with neutron source, triggering on phonon channels, -2.5 V DIB1, +2.5 V DIB2, 3.2768 ms trace length, 400 μs trigger delay: data series 07110604_2234 onwards. |