| 1123 | Stopped DAQ while cooling down after flashing. |
Data collected in the last run
3.7G 07110530_1900
3.7G 07110530_1948
3.7G 07110530_2035
3.5G 07110530_2123
3.6G 07110530_2208
3.7G 07110530_2254
3.7G 07110530_2341
3.7G 07110531_0028
3.7G 07110531_0115
3.7G 07110531_0202
3.7G 07110531_0250
3.7G 07110531_0337
3.7G 07110531_0424
3.7G 07110531_0512
3.7G 07110531_0559
3.7G 07110531_0647
3.7G 07110531_0734
3.7G 07110531_0821
3.7G 07110531_0908
3.7G 07110531_0955
3.7G 07110531_1043
| 1900 | Fill N2 cold trap full. |
| 1948 | Checked SQUIDs, no trapped flux. Found DCRC30 phA was actually using the settings for DCRC33 phA. Corrected it and saved the settings in oldDAQSettings_20110531.xml. Timing link and triggers are all set. |
| 2205 | Start taking Cf-252 data 07110531_2205_F0001. -2.5V on DIB1, 2.5V on DIB2. Trigger on 8 phonon channels. 30min data, 0.1min flashing, 15min cooldown. DAQ settings OldDAQSettings_20110531_2204.xml. Trigger threshold 18ADC bin. There seems many pileips. Found LED connection configuration can bring in noise. wr 30 800 seemed to get rid of the noise. The Cf-252 source is 17.5“ from the mu-shield. |