| 0100 | Fill N2 cold trap full. |
| | Start flashing detector. |
| 1050 | Filled LN2 Cold Trap. |
| 1340 | Start LHe transfer. Helium liquid level 3.0%. |
| 1415 | End LHe transfer. Helium level 94.1% . |
| 1500 | Stop flashing detector. |
| 1750 | Tower temperature a bit high at 66 mK. Pressure in P1 a bit low at 0.07 mbar - should be around 0.19 mbar at 68 mK. Increase still heater power to 3.0 mW. |
| 1810 | Tower temperature still a bit high at 63 mK. Pressure in P1 still a bit low at 0.085 mbar. Increase still heater power to 4.0 mW. |
| 1825 | Tower temperature still a bit high at 63 mK. It seems as though we have lost some of our cooling power. |
| | Start data series regardless: 07110511_1821: -2.5 V bias DIB1; +2.5 V bias DIB2; Ba-133 source 13” below top of mu-metal shield; 40 ADC bins trigger threshold on all charge channels; charge-only. |
| 1930 | Stop data series 07110511_1821 on file 14. |
| | Flash detector for 15 s using Berkeley's settings: LEDs 1 & 2 on both DIBs; 0.8 mA current; 50 μs on; 2.5 ms off. Tower temperature increases to 450 mK. |
| 1940 | DCRC30 started emitting a high pitch noise as soon as flashing started. Disconnecting it solved the problem. |
| 2025 | Tower temperature stuck at 80 mK, and pressures in G1 & G2 are fairly high at over 250 mbar. Decrease still heater power to 2.0 mW. |
| 2040 | Doesn't seem to be helping. Turn still heater off. |
| 2050 | Tower temperature falls very slowly, but it appears to be falling, which is an improvement. |
| | Fill N2 cold trap full. |