05/08/2011

1435 Tower temperature reads “vmix ovl”
1515 Filled LN2 cold trap.
1637 1K pot was running at 1.33K. Open the needle valve a little bit more to set temperature to ~1.5K. 1K pot temperature went to 1.66K suddenly and then slowly dropping. Tower Temperature 54.5mK.
1648 Power up DCRC boards, temperature went up from 53.1mK, then dropped.
1800 Start flashing detector.
2000 Stop flashing detector.
2100 Tower temperature still high at around 400 mK. Pressures in G1 & G2 around 60 mbar: it looks as though there is very little circulation.
Resistance of temperature sensor in still is a little higher than it should be at over 5 kΩ. Perhaps it is time to turn on the still heater.
Still heater turned on; power 1 mW.
2130 Tower temperature has fallen to 60 mK. Looks like still heater did the trick.
Resistance of still thermometer is a little lower than usual at 4.4 kΩ (should be around 4.6 kΩ). This is probably alright.
Pressures in G1 & G2 slightly higher at 70-80 mbar, but still a little low. As long as the base temperature hovers around this level, it might be better not to fix what isn't broken.
2230 Tower temperature has stabilised at 54 mK.
2240 Start data series: 07110508_2236; +3 V bias DIB1; -3 V bias DIB2; Ba-133 source 13” below top of mu-metal shield; 40 ADC bins trigger threshold on all charge channels; charge-only.