| 1053 | Flash detector |
| 1250 | Fill N2 cold trap full. |
| 1730 | Some issues with reducing the iZIPv4 data. Been working with Lauren to sort them out. Currently ongoing. |
| 1800 | Looks like problem was in the DAQ code. Jeter sent a new version. Have downloaded it to the laboratory computer. Will try it out later tonight. |
| 1920 | Have been trying to reduce data series 07110502_0044, but encounter the error: NoiseBuilder::ConstructMinMaxDistrubution ERROR in ordering of MapOfZipPulses or MapofNoisePulses!noise data is 193, and pulse data is 4001006 when running NoiseGen. Not sure what this means, but no such error crops up when attempting to generate a filterfile from data series 07110428_1946, collected under the same conditions. Are these files corrupted? |
| | Perhaps this isn't worth investigating in great detail at the moment seeing as we have received a new version of the DAQ. Time to test it out. |
| | Delete files taken from 2011/04/27 onwards on DAQ computer to make space. |
| | Stop flashing detector. |
| 2010 | Start data series with new DAQ: 07110502_2009; charge-only; trigger set at 30 ADC bins for each charge channel. |
| 2105 | New version of the DAQ seems to have done the trick. Ran NoiseGen and BatRoot on the first file successfully. |
| 2120 | Stop taking data. Start flashing detector. |
| 2215 | Stop flashing detector. |
| 2300 | Start taking data with Ba-133 source, series: 07110502_2259; charge-only; trigger set at 30 ADC bins for each charge channel. |
| 2350 | Fill N2 cold trap full. |