UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20110501

2011/05/01

0015 Stop taking data.
Flash detector overnight.
Fill N2 cold trap full.
1710 Fill N2 cold trap full.
1735 Stop flashing detector.
1750 Start LHe transfer.
1830 Stop LHe transfer.

Neutralisation

2300 Look at data series 07110428_1946. Histogram of Qsum shown in Figure 1 and scatter plot of Qsum as a function of time in Figure 2.
Figure 1: Qsum after low energy, Qpart & chi2 cuts.
Figure 2: Qsum after low energy, Qpart & chi2 cuts as a function of time.
The 60 keV line is not visible in Figure 2.
The threshold of 40 ADC bins (approximately 2.6 mV) might be too high and cuts out a number of 60 keV events. The relatively small number of events in the 60 keV peak in Figure 1 seems to indicate this. However, charge signals from 60 keV events should be around 3 mV in amplitude, and should satisfy the trigger threshold. Perhaps neutralisation degrades extremely quickly under bias, and the signals drop below threshold.
We could try a Ba-133 run to see if we get any joy.