| 1230 | Start circulation. |
| 1245 | Fill N2 cold trap full. |
| 1345 | Start flashing detector. LEDs 1 & 2 on both DIBs; 0.6 mA current, 100 μs on, 2.5 ms off. Tower temperature rises from 80 mK to 850 mK. |
| 1445 | Tower temperature has increased to >2 K. Not sure why. Temperature monitor indicates that temperature remained at around 850 mK for some time before rising to this level. Stop flashing detector to see if this fixes things. |
| 1505 | Improves the situation slightly, but tower temperature is still too high at 900 mK. |
| | Open needle valve to 1 K pot to allow more helium in. Temperature of 1 K pot rises from 1.45 K to 1.75 K. Tower temperature has now dropped below 300 mK. |
| 1515 | Close needle valve such that temperature of 1 K pot falls to 1.6 K. Tower temperature has dropped to 140 mK. |
| 1600 | Start LHe transfer. |
| 1610 | Tower temperature rose to 103 mK, and then dropped as transfer started. Currently at 77.5 mK and falling pretty rapidly. |
| 1650 | Stop LHe transfer. Helium liquid level 93.7%. |
| | Tower temperature 66.3 mK, falling slowly. |
| 1755 | Start flashing detector. LEDs 1 & 2 on both DIBs; 0.6 mA current, 100 μs on, 2.5 ms off. Tower temperature rises from 64 mK to 850 mK. |
| 1815 | Stop flashing detectors. |
| 2200 | Jeter sent 3 versions of the DAQ code today. Both have some issues. Trigger settings Qi 40 ADC bins and Qo 27 ADC bins on both DIBs. |
| | Version 2 collects random triggers, but stops shortly afterwards and spits out a 'Memory is Full' error after a few seconds. |
| | Version 3 does not get this far, and produces an error with the global trigger queue. |