2011-04-25
| 0948 | Filled N2 cold trap. MC X12004 88.5K. Temperature dropped 20K in the past 4 hours. IVC pressure 0.54mBar. The temperature dropping rate seems a bit slow. But the exchange gas pressure should be good enough, same as in the last run(?). |
|---|---|
| 1215 | Switch from cold trap 2 to cold trap 1 to facilitate cleaning of cold trap 2. |
| Mixing chamber temperature 83.0 K. | |
| 1525 | Pump out N2 exchange gas. |
| Mixing chamber temperature 79.6 K. | |
| 1625 | Stop evacuating IVC at pressure 4E-6 mbar. Background He leak rate 3.6E-9 mbar l/s, rising very slowly. |
| Start blowing out LN2. | |
| 1700 | Fill N2 cold trap full. |
| 1740 | LN2 out of main dewar. He leak rate rose steadily to 1.2E-8 mbar l/s: a bit high, but no spike so no significant leak. |
Electrical Check at 77 K
| 1755 | Perform electrical check. Mixing chamber temperature 78.9 K. |
|---|
| Detector: | G42 DIB1 | BH: | Stripline: | 20 | ||||
| SC: | ZB1-1-9 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2011-04-23 | Date | 2011-04-25 | |||
| SQUET: | ZC70 | Who | HC & JZ | Who | HC & JZ | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9 | OK | 1 - 2 k | 1.96 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.7 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.7 | OK | 1.82 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.3 | OK | 2.09 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29.4 | OK | 9 - 13 k | 15 | OK |
| Feedback B/Ret | 2 | 18 | 25.6 | OK | 9.2 | OK | ||
| Feedback C/Ret | 41 | 25 | 28.6 | OK | 4.2 | OK | ||
| Feedback D/Ret | 22 | 6 | 34.8 | OK | 17.4 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 79.5,81.2,79.2,80.1,80.1 | OK | 30 - 40 | 43,44,42,43,43 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 76,76,79,76 | OK | 37,37,40,37 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 78, 77, 69, 77 | OK | 41,41,33,40 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 81, 80, 80, 80 | OK | 44,43,43,43 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06 | OK | 1.0 - 1.1 k | 1.04 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 39MΩ | OK | < 1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 39MΩ | OK | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43.5MΩ, 43.5Ohm | OK | 6 - 8 nS | 50MΩ, 50MΩ | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 43.6MΩ, 43.6MΩ | OK | 50MΩ, 50MΩ | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.7 | OK | 0.9 - 1.0 V | 1.084 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.13 | – | 1.2 - 1.5 V | 1.411 | – |
| LED 2 | 26 | chassis | 1.12 | – | 1.403 | – | ||
| Detector: | G42 DIB2 | BH: | Stripline: | 19 | ||||
| SC: | ZB1-1-6 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2011-04-23 | Date | 2011-04-25 | |||
| SQUET: | ZC23 | Who | HC & JZ | Who | HC & JZ | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.63 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 1.99 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.73 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.58 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19 | OK | 9 - 13 k | 7.3 | OK |
| Feedback B/Ret | 2 | 18 | 21 | OK | 8.2 | OK | ||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.6 | OK | ||
| Feedback D/Ret | 22 | 6 | 18 | OK | 8.1 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 85, 87, 85, 86, 86 | OK | 30 - 40 | 44,46,44,45,46 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 86, 86, 89, 86 | OK | 45,44,47,44 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 86, 85, 75, 85 | OK | 45,45,34,44 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 85, 84, 84, 84 | OK | 45,44,44,44 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.069 | OK | 1.0 - 1.1 k | 1.044 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 38MΩ | OK | < 1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 38MΩ | OK | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43MΩ, 43MΩ | OK | 6 - 8 nS | 49MΩ, 49MΩ | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 43MΩ, 43MΩ | OK | 49MΩ, 49MΩ | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.721 | OK | 0.9 - 1.0 V | 1.091 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.138 | – | 1.2 - 1.5 V | 1.417 | – |
| LED 2 | 26 | chassis | 1.128 | – | 1.407 | – | ||
| 1800 | Pump mixture into dump in preparation for throughput test. |
|---|---|
| Pump out main bath. | |
| 1815 | Background He leak rate 1.3E-8 mbar l/s. Start throughput test. |
| Throughput test yields rate of 43 mbar/min. Pump mixture into dump. | |
| No significant change in He leak rate: no leak from mixing chamber pipes to IVC. | |
| 1830 | Pump out helium transfer stick. |
| 1850 | Stop pumping helium transfer stick. |
| 1910 | Evacuate 1 K pot. |
| 1930 | Overpressurise main bath with He gas. Background leak rate 9.1E-9 mbar l/s. He leak rate drops a little to 8.9E-9 mbar l/s: no leak. |
| 1940 | Fill IVC with 0.12 mbar He-4 exchange gas. |
| 2000 | Start first LHe transfer. |
| 2053 | MX X12004 was 43K while He level > 22%. Hassan said he put in 0.12mbar exchange gas before transfer. This should be good. Maybe the transfer was too fast. IVC pressure 2.2e-2 mbar. Temperature still drops kind of linearly. |
| 2140 | Stop LHe transfer. Main bath 97.6% full. Mixing chamber temperature 10.7 K and falling fast. Slope is still fairly linear. |
| 2150 | Pump out He-4 exchange gas. Mixing chamber temperature 8.0 K. |
| 2220 | Fill N2 cold trap full. |
| Start cryocleaning mixture through cold trap 1. Valve 6 open 13.8%. Leave running overnight. | |
| Open valves 8 and 11b to pump out any remaining mixture in cold trap 2. Leave open overnight. |