2011/04/23
| 1230 | Fill N2 cold trap full. |
|---|
Tower Level Continuity Checks
| 1400 | Perform tower level continuity checks. |
|---|
| Detector | G42 | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-1-9 | date/who: | 2011-04-23 | HC |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 130 |
| 2 | 4 | QETA-B | 100 to 200 | 248 |
| 3 | 4 | QET B | 100 to 200 | 238 |
| 4 | 6 | QETB-C | 100 to 200 | 265 |
| 5 | 6 | QET C | 100 to 200 | 208 |
| 6 | 8 | QETC-D | 100 to 200 | 264 |
| 7 | 8 | QET D | 100 to 200 | 196 |
| 2 | 8 | QETA-D | 100 to 200 | 231 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 783k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 789k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.104 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.104 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.4 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.1 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.1 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 25 nS |
| Detector | G42 | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-1-6 | date/who: | 2011-04-23 | HC |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 | 135 |
| 2 | 4 | QETA-B | 100 to 200 | 258 |
| 3 | 4 | QET B | 100 to 200 | 252 |
| 4 | 6 | QETB-C | 100 to 200 | 276 |
| 5 | 6 | QET C | 100 to 200 | 210 |
| 6 | 8 | QETC-D | 100 to 200 | 273 |
| 7 | 8 | QET D | 100 to 200 | 205 |
| 2 | 8 | QETA-D | 100 to 200 | 241 |
| chassis | 9 | LED1 Rforward | 8M to 9M | 789k |
| chassis | 10 | LED2 Rforward | 8M to 9M | 794k |
| chassis | 9 | LED1 Vforward | 1.0 V to 1.1 V | 1.107 V |
| chassis | 10 | LED2 Vforward | 1.0 V to 1.1 V | 1.108 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.4 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.1 nS |
| 1 | 11 | Qo Rbias | 24 nS to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 nS to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 nS to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.1 nS |
| 15 | 16 | Qi Rfeedback | 24 nS to 26 nS | 26 nS |
Berkeley Result:
| 2130 | Finish tower installation. |
|---|
Electrical check at 300 K
| Detector: | G42 DIB1 | Stripline: | 20 | ||
| SC: | ZB1-1-9 | 300 K | |||
|---|---|---|---|---|---|
| Face: | 5 | Date | 2011-04-23 | ||
| SQUET: | ZC70 | Who | HC & JZ | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | |
| SQUID bias C/Ret | 8 | 24 | 4.7 | OK | |
| SQUID bias D/Ret | 23 | 39 | 5.3 | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29.4 | OK |
| Feedback B/Ret | 2 | 18 | 25.6 | OK | |
| Feedback C/Ret | 41 | 25 | 28.6 | OK | |
| Feedback D/Ret | 22 | 6 | 34.8 | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 79.5,81.2,79.2,80.1,80.1 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 76,76,79,76 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 78,77,69,77 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 81,80,80,80 | OK | |
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.060 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 39 MΩ | OK | |
| Qo bias/QET Ret | 11 | 39 MΩ | OK | ||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43.5 MΩ, 43.5 MΩ | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 43.6 MΩ, 43.6 MΩ | OK | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | |
| Qi source | 31 | chassis | – | OK | |
| Qo source | 27 | chassis | – | OK | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.7 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.13 | – |
| LED 2 | 26 | chassis | 1.12 | – | |
| Detector: | G42 DIB2 | Stripline: | 19 | ||
| SC: | ZB1-1-6 | 300 K | |||
|---|---|---|---|---|---|
| Face: | 2 | Date | 2011-04-23 | ||
| SQUET: | ZC23 | Who | HC & JZ | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK |
| SQUID bias B/Ret | 19 | 34 | 5.0 | OK | |
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | |
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19 | OK |
| Feedback B/Ret | 2 | 18 | 21 | OK | |
| Feedback C/Ret | 41 | 25 | 19 | OK | |
| Feedback D/Ret | 22 | 6 | 18 | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 85,87,85,86,86 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 86,86,89,86 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 86,85,75,85 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 85,84,84,84 | OK | |
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.069 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 38 MΩ | OK | |
| Qo bias/QET Ret | 11 | 38 MΩ | OK | ||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 43 MΩ, 43 MΩ | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 43 MΩ, 43 MΩ | OK | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | |
| Qi source | 31 | chassis | – | OK | |
| Qo source | 27 | chassis | – | OK | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.721 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.138 | – |
| LED 2 | 26 | chassis | 1.128 | – | |
| 2300 | Evacuate IVC overnight. |
|---|