2011/04/11

1015 Automatic data taking is finished. The fridge worked fine the whole night. We can continue to take data without problems.

1042 Filled LN2 Cold Trap.
1142 Started data 07110411_1142_F0001. Calibrated -4V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 60.5mK.
2028 The timing link of the data took after +6V bias sweep wasn't set correctly and a large fraction of the data was noise. I modified the SCDMS_DetectorReadout.vi to have a look at the baseline standard deviation. So I can set my trigger level as a number of sigmas away.
2031 Q1 baseline std: 15.20ADC. Q3 baseline std: 8.05ADC. Q2 baseline std: 11.20ADC. Q4 baseline std: 9.50ADC.
I will set trigger level as 3sigma. Then they are Q1=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D).
2049 Checked timing link from the readout vi: OK. Checked trigger settings from terminal: OK.
2053 Started data 07110411_2053_F0001. Calibrated -8V 1hour neutralization data after 20min flashing. Trigger: Q1=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D). I can see decent synchronized pulses on the screen. The run should be good.
Will take -7V to -1V in -1V step and -(1,2,3,4)V and -(4,3,2,1)V data overnight.
2104 Filled N2 cold trap.
2134 Stop the DAQ after 15min data and 20min flashing. Had a look at the data. Timing link and trigger levels both were OK. Q4 had the most events. Q1 and Q3 didn't have so many.

-7V 1 hour neutralization

Data:

07110411_0044

Script

Plots

-8V 1 hour neutralization

Data:

07110411_0245

Script

Plots

-1V 1 hour neutralization

Data:

07110411_0445

Script

Plots

-(1,3,2,4)V 1 hour neutralization

Data:

07110411_0646

Script

Plots

-(4,2,3,1)V 1 hour neutralization

Data:

07110411_0846

Script

Plots

-4V 15min

Data:

07110411_1142
07110411_1304
07110411_1420
07110411_1536
07110411_1652

Script

Plots

-8V trigger check

Data:

07110411_2053

Script

Plots