Table of Contents
2011/04/10
| 0039 | Stop DAQ after 1hour run. I don't think the fridge can take any heat from flashing. I'll leave it like this. Disable charge polarity flipping. |
|---|---|
| 1004 | Without flashing, the fridge was running OK except 1K pot was a bit cold. |
| 1009 | Open 1K pot a bit more. Pot temperature went up to 1.7 and drops slowly. I'll try to flash the detector. |
|---|---|
| 1057 | Flashed detector for 20min. Everything went OK. Stop the DAQ. |
| 1205 | Start to take data 07110410_1205_F0001. Calibrated -2V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. Tower T 61.7mK. |
| I forgot to stop the run after flashing. One more data set 07110410_1406 was taken for -2V bias. | |
| 1430 | Filled N2 cold trap. |
| 07110410_1611: 1 hour data at -3V after 20min flashing. 07110410_1814: 1 hour data at -4V after 20min flashing. | |
| 2043 | Start -5V 1 hour run. This is the first run of automatic data taking. Will take -6, -7, -8, -1234, -4321 biases. |
| 2051 | Filled N2 cold trap. Turned off Helium level meter to avoid the labview error. |
-2V 1 hour neutralization
Data:
07110410_1205
07110410_1406
The cuts are defined as
- cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
- cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq
- cE1 = 15 < Q1;
- cE2 = 15 < Q2;
- cE3 = 10 < Q3;
- cE4 = 20 < Q4;
- cE = 50 < Qsum.