2011/04/09
| 0024 | Stop flashing. |
|---|---|
| 0104 | Start 07110409_0104_F0001. DAQ flashed the detector again. Need to wait to cool down. |
| 0154 | Start 07110409_0154_F0001. Calibrated +6V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 60.8mK. |
| 0949 | 1K pot ran away at about 7am in the morning. Stop taking data at 07110409_0925. Stop flashing started by DAQ. Pump mixture into dump. The data sets up to 07110409_0655 are good. |
| 0953 | Close V4A, stop 1K pot. See if warming up the 1K pot to 4K would help. |
|---|---|
| 1031 | Filled N2 cold trap. |
| 1115 | Open V4A to start 1K pot. P2 drops quickly. The blockage is still there. Open the 1K pot needle value a little bit more. 1K pot temperature and p2 both increase. The fridge can still operate in this mode. |
| 1121 | Close V5 and V6. Open V4 and V3 to condense mixture. G1=206mbar, G2=361mbar. P2=20mbar, 1Kpot 1.646K. It maybe a better idea to let mixture run through N2 cold trap when condensing. |
| 1136 | The 1K pot still running down. But I'll start circulation anyway. |
| 1138 | 1K pot is very sensitive to the change in circulation line. I think the pot doesn't have much liquid helium in it. This may be due to a blockage in the he intake line. |
| 1546 | Tower temperature 60.7mK. 1K pot still keeps running down but the fridge otherwise is running OK. I'll take more data. |
| 1547 | Start data 07110409_1547_F0001. Calibrated +8V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 60.7mK. |
| 1556 | Stop the run. Q2 and Q4 have a baseline around -2V. Q2 actually railed. After setting bias to 0V. Q2 and Q4 still behave the same. I'll flash the detector for a while and see if it'll be fixed. |
| 1930 | Begun Helium transfer. |
| 2028 | Finished Helium transfer. With -4V bias Q2 and Q4 still have -2V offset. Cycle power of two boards. |
| 2030 | After cycle the power, Q2 and Q4 come back. Q1 and Q3 are still fine. |
| 2100 | Flash detector. Tower temperature went up to 547mK. Looked fine. |
| 2120 | Stop flashing the detector. |
| 2128 | Created DAQ setting files with different biases. |
| 2133 | Fill N2 cold trap. |
| 2335 | Start taking data 07110409_2335_F0001. Calibrated -1V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. Tower T 61.5mK. The polarity was flipped. |
| 2343 | 1K pot 1.345K, keeps dropping |
Calibrated +6V bias
Data:
07110409_0154
07110409_0309
07110409_0425
07110409_0540
07110409_0655
The cuts are defined as
- cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
- cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq
- cE1 = 15 < Q1;
- cE2 = 15 < Q2;
- cE3 = 10 < Q3;
- cE4 = 20 < Q4;
- cE = 50 < Qsum.
Comments:
Plots
-1V 1 hour neutralization
Data:
07110409_2335
I don't know where the gap in the event time is from. This maybe due to computer time error or labview.