UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20110408

2011/04/08

0830 Temperature looked OK in the past 12 hours.
0833 Stop +2V data at 07110408_0735. Load +4V bias into DAQ.
0844 Filled N2 cold trap.
0859 Start to take data 07110408_0859_F0001. Calibrated +4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 62.8mK.
1016 1K pot started to run down from early morning. Tune it back to ~1.5K

1458 1K pot kept running down. There might be a blockage in the pumping line.
1848 Stop +4V bias sweep.
2257 Modified SCDMS_GetTraceData.vi to flip the sign of trace traces. Created +6V bias setting file. Changed trigger settings to 60, 36, 30, 30 ADC bins based on the +1V data.
2259 Flash detector.
2302 Filled N2 cold trap.
2309 Set 1K pot to 1.6K. Hopefully it won't drop too low soon.

Calibrated +4V bias

Data:

07110408_0859
07110408_1016
07110408_1131
07110408_1247
07110408_1403
07110408_1520
07110408_1635
07110408_1750

Scipt

The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq
  • cE1 = 15 < Q1;
  • cE2 = 15 < Q2;
  • cE3 = 10 < Q3;
  • cE4 = 20 < Q4;
  • cE = 50 < Qsum.

Comments:

The trigger threshold is too low for this series. Most of the events are noise.