2011/4/5
| 0934 | Filled N2 cold trap. |
|---|---|
| 1454 | Stop the DAQ after 07110405_1325. Flash the detector. The series 07110405_1456 has too few events even to make a filterfile. |
| 1954 | Stop flashing detector. |
| Filled N2 cold trap. | |
| 2059 | Start taking data 07110405_2059_F0001. Calibrated -6V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. |
Calibrated -6V bias
Data:
07110404_2119
07110404_2320
07110405_0121
07110405_0321
07110405_0522
07110405_0724
07110405_0924
07110405_1125
07110405_1325
The cuts are defined as
- cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
- cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq
- cE1 = 15 < Q1;
- cE2 = 15 < Q2;
- cE3 = 10 < Q3;
- cE4 = 20 < Q4;
- cE = 50 < Qsum.
Comments:
- Not so many events on Q1 and Q3. The change is not unexpected compared with uncalibrated -6V data.
- Cross talk is strange.