UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20110402

2011/04/02

0009 Start taking data 07110402_0009_F0001. Calibrated -4V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 30min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF.
Trigger rate was too low. Stop the run and reset timing link.
0013 Start 07110402_0013_F0001
0935 The 'Not enough memory to complete this operation' error occurred again close to the end of the first file. Not sure why. After clicking OK the VI continued to run, but another error window popped up. Some screen shots are here
0953 Flashing started by the DAQ vi. I'll stop the vi and flash for a while.
1235 Fill N2 cold trap full.
1643 Created DCRC_ZeroCharge.vi to set charge channel to the calibrated zero. Turn on LED1 on DIB1 to continue flashing. Tower temperature ~560mK.
1715 Stop flashing detector
1727 Not all the LEDs were turned off. Turn them all off.
1805 Stared taking data 07110402_1805_F0001. Calibrated -4V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 30min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF.

Calibrated -4V bias 07110402_1805

Series 07110402_1805, 07110402_1936, 07110402_2106, 07110402_2236, 07110403_0007, 07110403_0137, 07110403_0307, 07110403_0437, 07110403_0608, 07110403_0738, 07110403_0908, 07110403_1039

Calibrated -4V on Q1-4. The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.
chisq cuts of Q1-Q4
Fig. 1.1
Fig. 1.2
Fig. 1.3
Fig. 1.4
Neutralization of Q1-Q4
Fig. 1.5
Fig. 1.6
Fig. 1.7
Fig. 1.8
Fig. 1.9
Charge energy spectrum
Fig. 1.10
Fig. 1.11

Headline

Data:

07110402_1805
07110402_1936
07110402_2106
07110402_2236
07110403_0007
07110403_0137
07110403_0307
07110403_0437
07110403_0608
07110403_0738
07110403_0908
07110403_1039

Script

Plots