2011/04/01

0859 Stop the run at 07110401_0819_F0006. The system is cooling down after a 20min flashing. Stop the DAQ.
0941 Start taking data 07110401_0941_F0001. -1V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~62.4mK.
1003 Filled N2 cold trap
1820 Begin pumping on helium transfer stick.
1901 Pumped helium transfer stick to 4.1e-3 mbar
1930 LED1 on ZIB1 didn't work a moment ago. Cycle the power of DCRC30 and it worked. Flashing the detector. Zero biases using true bias values.
1950 Begun helium transfer.
2035 Finished helium transfer.
2050 Filled N2 cold trap.
2318 Stop flashing.

-1V bias 07110401_0941

Series 07110401_0941 07110401_1056 07110401_1212 07110401_1327

-1V on Q1-4. The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.

Data:

07110401_0941
07110401_1056
07110401_1212
07110401_1327

Script

Plots