2011/03/31

0006 Started taking data 07110331_0006_F0001. -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~62mK.
0919 Labview stopped at File 9 of series 07110331_0607 because 'not enough memory'. There might be a memory leak. Have to kill Labview to end it.
0930 Filled N2 cold trap. Tower T=61.7mK.
1059 Started flashing the detector. All 3 LED's used. 0.6mA, 100us ON, 2500us OFF. Tower T rose up to 540mK.
1239 Stop flashing detector.
1329 Start to take data 07110331_1329_F0001 after 40min cooldown. -6V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~60.5mK.
1413 A power glitch occurred during the cooldown after flashing. The temperature went up. Both pumps stops. I started them right away. This caused about 7min delay in the cooldown. I'll pause the vi for additional 7min.
1830 Start to take data 07110331_1830_F0001. -8V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~65mK.
1936 Filled N2 cold trap.
2320 Changed the DCRC setup from indicator to control in DCRCv3MultiDAQ.vi and the settings can be changed from the front panel.
2334 Start to take data 07110331_2334_F0001. -2V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~63.4mK.

-4V 1 hour Neutralization

Data: Series 07110331_0006, 0206, 0406, 0607

Don't know what happened in the seires _0006. We decided to take 15min of data and use the same flashing scheme. Take 4 datasets for each bias.

The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.
chisq cuts of Q1-Q4
Fig. 1.1
Fig. 1.2
Fig. 1.3
Fig. 1.4
Neutralization of Q1-Q4
Fig. 1.5
Fig. 1.6
Fig. 1.7
Fig. 1.8
Fig. 1.9
Charge energy spectrum
Fig. 1.10
Fig. 1.11

-4V neutralization again

Data:

07110331_0006
07110331_0206
07110331_0406
07110331_0607

Script

Plots

-6V 15min bias sweep

Data: Series 07110331_1329, 1444, 1559, 1714

-6V on Q1-4. The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.

Data:

07110331_1329
07110331_1444
07110331_1559
07110331_1714

Script

Plots

-8V 15min bias sweep

Series 07110331_1830, 1945, 2100, 2216

-8V on Q1-4. The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.

Data:

07110331_1830
07110331_1945
07110331_2100
07110331_2216

Script

Plots

-2V 15min bias sweep

07110331_2334, 07110401_0049, 07110401_0204, 07110401_0319, 07110401_0434, 07110401_0549, 07110401_0704, 07110401_0819,

-2V on Q1-4. The cuts are defined as

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 50<Qsum.

Data:

07110331_2334
07110401_0049
07110401_0204
07110401_0319
07110401_0434
07110401_0549
07110401_0704
07110401_0819

Script

Plots