| 0753 | Stop the run at 07110329_2325_F0162. At about 06:14 in the morning. A temperature glitch happened and then temperature drops. Tower temperature 43.1mK. |
| 0758 | Start to flash detector. |
| 0942 | Tower temperature went up to 1.4K. Stop flashing. |
| 0959 | G1=766mbar, G2==696mbar. The mixing process must stopped. Need to recondensing again. I think the fridge entered one shot mode after the temperature glitch. There might be a blockage in the circulation line. |
| 1002 | Closing V6. P1=0.032mbar. |
| 1005 | V6 is closed. P1=0.7mbar. G1=766mbar, G2=696mbar |
| 1006 | Open v3 to let mixture condense from both ends of fridge. G1=126mbar. G2=378mbar. V12A=100%. |
| 1008 | G2 and G1 keep dropping. Tower T=1.77K. |
| 1016 | G1=57mbar, G2=51mbar and both plateaued. The condensing is done. Now close V3 and open V6 to 20% to start circulation. |
| | G1 and G2 shoot above 600mbar. Close V6 to 13%. |
| 1021 | G1 = 667mbar and G2=604mbar. G1 and G2 don't change when V3 is closed. It looks like there is a blockage in the condenser line. It is probably the He cold trap. |
| 1105 | Filled N2 cold trap. |
| 1106 | Closed condenser speedivalve. |
| 1107 | Close V13A, open V9, V5 to pump the N2 and He cold traps. Close V6 all the way. |
| 1449 | Open V14 to let the mixture expand if to the dump if they would. G1=2.5mbar and almost plateaued. |
| 1654 | G1=2.3mbar. G2=30mbar, P1=75mbar. Close V12A. |
| 1751 | G1 = 1.0mabr, G2=31.3mbar, P1=1000mbar. Can not pump G1 lower. Close V1, G1 stays about the same. Open V2 to pump mixture out of the lines. |
| 1811 | G1=0.8mbar. Close V2. G1 fluctuates around 1mbar. |
| 1812 | Close V5, G1 increases to 4mbar. Open V5, G1 drops to 1mbar. Open V5, V1, pump G1 to 1mbar and then close V5, G1 rise to 2.3mbar. |
| 1822 | Pull out the He cold trap and G1 went up to 50 mbar. |
| 1823 | Switch He4 pump to exhaust. Close V4A, Open V5A, V2A V7, V2 to pump the He cold trap. |
| 1840 | Heat the He trap to above 100C and continue pumping. Pull out N2 cold trap 1, G1 increases. Open V12A. Heat N2 cold trap 1. |
| 1909 | There might a leak from the N2 cold trap 1 safety valve. Tighten it. |
| 1942 | G1 = 1.3mbar and is nearly plateau. Close V2, V7, V2A, V4A. Open V4A to start 1K pot. Switch He4 pump to He recovery. |
| 1945 | Put N2 cold trap 1 into liquid N2 dewar. |
| 2009 | Insert He cold trap into the bath. Open condenser speedivalve, start circulate mixture. The blockage seems to be removed. |
| 2032 | Begun Helium transfer. |
| 2033 | Close V14 at P1=1.3mbar. G1=82.5mbar, G2=94.7mbar. |
| 2114 | Finished Helium Transfer |
| 2125 | Tower T=60mK. The fridge is back to business. Start flashing the detector. |
| 2323 | Stop flashing detector. |