2011/03/27
| 0959 | Stop flashing detector |
|---|---|
| 1028 | Filled N2 cold trap. |
| 1127 | Trigger bits and thresholds have to be set in 'SCDMS_DetectorReadout.vi'. The MultiDAQ vi only matches trigger bits but it doesn't configure the trigger setting to the board. |
| 1326 | Experiment different ways to neutralize Q1 and Q3. Take files with 500 randoms + 500 events to check if Q1 and Q3 are responsive. |
| Start to take data 07110327_1326_F0001, -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. Tower T=57.5mK. | |
| 1358 | Q1 and Q3 are still not very responsive in 07110327_1326 even after overnight baking. The event fraction for Q1-4 are [0.0130, 0.6089, 0.0569, 0.3212]. I'll try flashing the detector with positive bias to force the surface to neutralize. |
| 1407 | Flash the detector with +4V on Q1-4. |
| 1516 | Set bias to 0V on Q1-4. Keep flashing. |
| 1533 | Stop flashing detector. |
| 1607 | Start taking data 07110327_1607_F0001 after flashing. -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. |
| The trigger rate of the run was very low and the two boards didn't seem to be synchronized. Stop the run. | |
| 1612 | Took new series 07110327_1612_F0001 after re-setting the timing link. Trigger rate was recovered. Q1&3 looked more responsive on the screen. |
| 1642 | Stop the run after 30min. |
| 1750 | Filled N2 cold trap. |
After flashing with positive bias
After flashing with the charge electrodes positively biased and then biased at 0V, I took some data to check the neutralization effects. The primary events fractions of Q1-Q4 are [0.0720, 0.5503, 0.0294, 0.3483]. Q1 and Q3 were still weak after flashing. The energy in the figures is not calibrated. The cuts are defined as
- cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
- cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
- cE = 45<Qsum.