2011/03/26

1103 Stop flashing detector
1126 Filled N2 cold trap
1130 Unbridge the analog grounds of DCRC. The noise seemed to be smaller.
1152 Start taking data 07110326_1152_F0001. -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DCRC30 on DIB2 (Q1&3), LED 1&2; DCRC33 on DIB1 (Q2&4), LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~60mK. DCRC30 on DIB2, master; DCRC33 on DIB1, slave.
1655 The series 07110326_1552 finished. DAQ started auto flashing. Stop DAQ to transfer He. Keep flashing.
1715 Begun Helium transfer.
1800 Stopped Helium transfer.
1918 Filled N2 cold trap

After overnight baking

Q1 and Q3 still didn't work well after overnight baking.

Cuts:

  • cQ# = max([Q1, Q2, Q3, Q4], [], 2)==#
  • cQCchisq = cQ1chisq & cQ2chisq & cQ3chisq & cQ4chisq;
  • cE = 45<Qsum.
chisq cuts of Q1-Q4
Neutralization of Q1-Q4
The figures below plot the primary events as a function of time. Q1 and Q3 have much fewer events than Q2 and Q4.
Charge energy spectrum