UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20110323

2011/03/23

0058 Start to take data 07110323_0058_F0001. -4V on Q1-4. Trigger: Q1=48, Q2=51, Q3=75, Q4=59, coincidence time 200us. Am-241. 1024 Q samples, 100us lookback time.
1400 Stop the run 07110323_0058
1438 Zap all squids on ZC70 (DCRC33). before take modulation curves.
1705 ZAP SQUET ZC23 (DCRC30).
1754 Modulation curves taken

<note> SQUET ZC23 and ZC70 SQUID modulation curves </note>

1818 Power off DCRC30 and remove it from ebox.
1836 Power up DCRC33 after removing DCRC30
1845 DCRC33 on DIB2, Q1 noise: 3mVpp, 1.2mVrms, base line is flat. Q3 noise: 2mVpp, 0.9mVrms, baseline wiggles.
Swap two boards → DCRC30 on DIB2
1924 Put DCRC33 on DIB1 (SL19) and have a look at the noise of both charge channels.
1938 Put DCRC30 on DIB2 (SL20) without powering it up. Have a look at the noise of DCRC33 on DIB1. The wiggling baseline came back again.
2107 Had a look at the charge noise with different DCRC connections.

<note> DCRC v3 & G101 Charge Noise </note>

2109 DCRC30 is connected to DIB2. DCRC33 unplugged. Flash the detector. Tower temperature went up to 440mK.
2119 Fill N2 cold trap.
2336 Stop flashing detector

-4V long neutralization

Data:

07110323_0058

Script

Plots