| 0022 | Begun new Ba-133 run 07110322_0020. Bias -4V on Q1-4. Trigger: Q1 OR Q2 OR Q3 OR Q4. Threshold: 64 ADC bins. |
| 0030 | Tower temperature at 63.3 mK. |
| 0922 | Stop the run 07110322_0020 at File 182 |
| 0935 | Flash the detectors and fill N2 cold trap full |
| 1004 | Begun Helium Transfer. |
| 1906 | EventTime is fixed. |
| 1907 | Unplugged the power the DCRC30 on zip2 (DIB1) and the large noise on Qo (Q3) of zip1 was still there. |
| 1909 | Unplugged the power DCRC33 on DIB2. |
| 1926 | Bridging the analog grounds of the two boards had no effect to the noise. RMS noise: Q1~0.8mV, Q2 and Q4 ~1mV, Q3~1.2mV. |
| 1947 | Measure RMS noise. Q1: 1.0 mv; Q2: 1.06 mV; Q3: 1.56 mV; Q4: 1.22mV. The trigger threshold can be set at Q1 = 48 ADC bins, Q2 = 51, Q3= 75, Q4 = 59. |
| 2007 | Flash detector |
| 2048 | Stop flashing |
| 2117 | Modified trigger setting vi so that I can set threshold on each channel independently. |
| 2136 | Fill N2 cold trap full |
| | Flash detector |